J&M TIDAS® L 900 NIR Near-Infrared Spectrometer

The TIDAS® L 900 NIR (a brand under the German Pharma Test Group, J&M) is a high-performance, integrated diode array near-infrared spectrometer designed for pharmaceutical quality control (QC/QA), research and development laboratories (R&D), raw material identification (Raw Material ID), blend uniformity, and process analysis (PAT). The device integrates a high-efficiency 7.5W halogen light source, a 256-pixel high-sensitivity InGaAs diode array detector, and an embedded touch-screen industrial computer system into a single unit, enabling non-destructive, rapid, and highly reproducible real-time acquisition of near-infrared spectral information in the 1100–2100 nm range.


I. Integrated independent touch architecture with built-in 10.4-inch touch screen, Win 10 IPC, 64GB SSD and TIDASDAQ 3 software; body size 47 × 37 cm, 12 kg, no external computer required for testing upon power-on.


II. The high-speed, non-moving optical system is equipped with a 7.5W halogen light source and an InGaAs 256-pixel diode array (1100 ~ 2100 nm), featuring a solid-state anti-vibration design and supporting millisecond-level full-spectrum real-time acquisition.


III. Excellent optical precision and signal-to-noise performance: Spectral resolution < 6 nm, wavelength accuracy ±0.5 nm, reproducibility < 0.01 nm, photometric accuracy ±10 mAU, SNR > 250:1, 16-bit conversion ensures high-precision analysis.


IV. Fiber Coupling and Compliance: Standard configuration includes SMA 905 fiber optic interface, 2 IN / 2 OUT digital I/O and Gigabit network; compliant with 21 CFR Part 11, ALCOA+, USP <856> and EP <2.2.40>, with original manufacturer IQ/OQ.



Equipment Introduction


  • Integrated Architecture & Standalone Touchscreen Control

    • Embedded industrial PC and intuitive interface : It features a built-in 10.4-inch high-resolution color touchscreen and an industrial-grade PC (IPC), pre-loaded with the manufacturer's TIDASDAQ 3 spectral analysis and data acquisition software, allowing for immediate testing without the need for an external computer.

    • Plug and play and compact size : The base measures only about 47 × 37 × 28 cm and weighs about 12 kg. It supports a wide voltage input of 85-265 VAC, making it ideal for use in quality control laboratories (QC/QA), research and development centers (R&D), or cleanroom inspection stations.

  • Solid-State High-Speed NIR Optical Engine (No Moving Parts)

    • Highly stable halogen light source and wide band coverage : Integrating a 7.5W tungsten halogen long-lasting light source with an InGaAs 256-pixel diode array detector , the spectral measurement range covers 1100–2100 nm (near-infrared first and second order overtone/combined frequency range).

    • All-solid-state shock resistance and ultra-fast acquisition : There are no mechanical spectrometers, interferometers or mirror moving parts inside, eliminating the risk of mechanical wear and wavelength drift; it has excellent shock resistance and supports millisecond-level (ms) full-spectrum high-speed synchronous acquisition.

  • Superior optical precision and signal-to-noise ratio (SNR) performance

    • High-precision optical specifications : spectral resolution < 6 nm (FWHM) , wavelength accuracy up to ±0.5 nm , wavelength reproducibility < 0.01 nm , absorbance/photometric accuracy up to ±10 mAU .

    • High signal-to-noise ratio and high dynamic range : Signal-to-noise ratio (SNR) > 250:1 , combined with a 16-bit A/D analog-to-digital converter , ensures that it can still output highly linear and reproducible spectral data under weak transmission or diffuse reflection signals, laying a solid foundation for chemometrics models.

  • Versatile Fiber Coupling, PAT & Compliance

    • Standard SMA 905 fiber optic interface : Supports standard SMA 905 fiber optic coupling and can be flexibly paired with solid diffuse reflection probes, liquid immersion transmission probes, non-contact optical heads, or flow cells.

    • Industrial-grade communication and PAT interoperability : Built-in 2 IN / 2 OUT digital I/O interface and Gigabit LAN Ethernet port, enabling seamless integration of process equipment (such as flow layers, mixers, and ingot presses) for real-time PAT interlock control.

    • Global pharmacopoeia and data integrity compliance : Fully compliant with FDA 21 CFR Part 11 , EU GMP Annex 11 , ALCOA+ , USP <856> (near-infrared spectroscopy) and EP <2.2.40> , standard electronic signature and audit trail, and provides original manufacturer's complete IQ/OQ validation documentation.

  • Core Applications

    • Raw Material ID : Supports non-destructive and rapid in-situ identification of APIs and admixtures (RMID) through transparent plastic bags or glass containers.

    • Blend Uniformity (BU) : Real-time monitoring of the powder mixing process and automatic determination of the endpoint.

    • Moisture Monitoring (LOD) for Drying and Granulation : Real-time online quantitative monitoring of dynamic moisture changes in the fluid layer or granulation process.

    • Tablet and Capsule Content Uniformity (CU) : A non-destructive, rapid screening method for the content and physical properties of active ingredients in finished or semi-finished products.




Product Features

Integrated, compact design; no external computer required.
High-speed, component-free optical architecture
Pharmaceutical Compliance
Process Analysis (PAT)
Expandability

Overall Specifications


  • Wavelength range : 1100 – 2100 nm

  • Spectral resolution : < 6 nm FWHM

  • Wavelength accuracy : ± 0.5 nm

  • Absorbance accuracy : ± 10 mAU

  • Wavelength reproducibility : < 0.1 nm

  • Signal-to-noise ratio (SNR) : 250:1

  • Built-in light source : 7.5W halogen lamp

  • Diode/Pixel Count : 256

  • Fiber Optic Connection Interface : SMA 905

  • Lighting connection interface : SMA 905

  • Analog/Digital Converter : 16-bit High-Resolution Converter

  • Digital input/output : Standard configuration 2 x IN / 2 x OUT

  • Communication interface : TCP/IP 10/100/1000 Mbit/s Ethernet

  • Embedded industrial computer : 10.4-inch color touchscreen, Windows 10 operating system, 64GB SSD solid-state drive.

  • Desktop space requirement : approximately 47 x 37 cm

  • Packaging and shipping dimensions : Approximately 80 x 60 x 60 cm (W x D x H)

  • Machine weight : Net weight approximately 12 kg / Gross weight approximately 18 kg

  • Power supply requirements : Wide voltage 85 – 265 VAC / 47 – 63 Hz

  • Standard Scope of Supply : TIDASDAQ 3 Spectral Analysis Software, Power Supply, Keyboard, Mouse, User Manual